High Frequency Mobile Surgical X-ray System AM-AR-M20E
- Home
- High Frequency Mobile Surgical X-ray System AM-AR-M20E
Personal Protective Equipment
Laboratory Equipment
Operating Theatre Equipment
Orthopedics: osteopathy, displacements, interlocking.
Surgery: foreign body removal, cardiac catheter, pacemaker implantation, interventional therapy, partial X-ray, local photography and other work.
- New integrated design, compact appearance, convenient operation.
- Unique portable remote control, more convenient operation.
- Colorful graphic LCD touch screen console interface, convenient operation, modern design.
- High quality HF-HV combined X-ray generator, greatly reducing X-ray irradiation.
- Toshiba image intensifier is used, which is stable and reliable in quality, with high-definition image display system.
- With KV, MA perspective auto tracking function, image brightness and definition are automatically set to the optimal function.
- Integrated digital acquisition and processing system makes operation more convenient.
1, New (with electric auxiliary support arm) C-arm main frame 1 set
- High voltage and high frequency X-ray generator and high frequency inverter power supply.
(5KW,40KHZ,120KV ) 1 set
3, Toshiba 9 inch image intensifier, 1 set
4 Megapixel Ultra Low Light Digital Cameras, 1 Set
5, Integrated digital acquisition and processing system 1 set
6, high thickness grids 1 set
7, adjustable electric collimator 1 set
8.24 inch LCD monitor, resolution: 1920*1080 1 set
9, 1 exclusive portable remote control set
PARAMETER | SPECIFICATION | |
Fluoroscopic Ability | Maximum nominal capacity | Tube current 4 mA, |
Tube voltage 120 kV | ||
Automatic fluoroscopy | Tube voltage: 40 kV ~ 120 kV Automatic adjustment | |
Tube current | 0.3mA ~ 4mA manually set | |
Fluoroscopia manual | Continuous tube voltage: 40 kV ~ 120 kV | |
Direct current tube | 0.3 mA ï½½ 4 mA | |
Wrist fluoroscopy | Continuous tube voltage 40 kV ~ 120 kV | |
Direct current tube | 0.3 mA ~ 30 mA | |
Photography capacity | Maximum nominal capacity | 5 kilowatts |
Combination of tube voltage and current
| 40kV × 49kV 1 × 180 mAs | |
50kV × 59kV 1 × 140 mAs | ||
60kV × 69kV 1 × 125 mAs | ||
70kV × 79kV 1 × 110 mAs | ||
80kV × 89kV 1 × 100 mAs | ||
90kV × 99kV 1 × 80 mAs | ||
100 kV ~ 109 kV 1 ~ 63 mAs | ||
110 kV ~ 120 kV 1 ~ 50 mAs | ||
Plate carrier size | 200 mm × 250 mm (8 ″ × 10 ″ ) o 250 mm × 300 m ( 10 ″× 12 ″) | |
X-ray tube | Special X-ray tube for high frequency | |
Fixed anode Double focus | Large focus: 1.5, small focus: 0.3 (advantage) | |
Inverter frequency: | 40 kHz | |
Thermal capacity: | 35KJ (47KHU) | |
Video system | Image intensifier | Image intensifier manufactured by TOSHIBA (9 '' , 6 '' , 4.5 '' ) |
Vidicón CCD | Ultra-low illumination megapixel CCD camera imported from Japan | |
Monitor | 23 inches, 1280 horizontal lines and 1024 vertical lines. | |
Bandwidth: | 48 MHz, frame/sec: 75 Grayscale: ≥ 11 Line pair resolution: ≥ 1.2 LP/mm | |
Integrated digital acquisition and processing system | Record: registration, medical history, work list Collection: Start collection; prepare video, restart, horizontal mirror, vertical mirror, window adjustment, magnifying glass, negative image Open silhouette, edge enhancement, recursive noise reduction Processing; four windows, nine windows, sharpness, horizontal mirror, vertical mirror, text annotation, length measurement report: save, preview, expert template Dicom Features: Dicom Browsing, Web Service | |
Structure | Steering wheel | ±90° revolution can freely change the movement direction of the unit. |
Ascending and descending range of the pillar | ≥ 400 mm | |
C-Arc | Forward and backward movement: 200 mm | |
Revolution around the horizontal axis | ±180° | |
Revolution around the vertical axis | ±15° (advantage), better than others ±12.5°, easier to take special positions (X-ray) | |
Orbital slippage | 120 ° (+90 °ï½ž 30 ° ) | |
Frame Monitor Rotation | ≥ 300 ° |