fbpx

Single/ ROII Double R.O Water Purification Machine AM-ROI

AM-ROI

Single/ ROII Double R.O Water Purification Machine AM-ROI

Thermal disinfection system

– Adopt heat disinfection system to disinfect RO membrane, RO water supply pipe, pipe for connecting with HD machine and port.

  1. No chemical residues
  2. Environmental protection
  3. No endotoxin remaining

Frequency converter control system

  1. Apply frequency converter and microelectronics technology to control AC motor power control system through changing motor frequency.
  2. Strictly control the water temperature between 85 and 95℃.

Online monitoring

– Online monitoring of chlorine and hardness

Bacteria filter

– Remove endotoxin in RO water by constant UF.

– The below items in the recycling RO water in the pipe can be removed.

  1. microorganism
  2. Suspicious particles
  3. biofilm fragment
  4. endotoxin

Water treatment system

– Sediment filter (with automatic rinsing device): Removal of particles and ferri ions from the water

– Carbon filter (With automatic rinsing device): Removal of chlorine and soluble organic matter.

– Softener filter (With automatic rinsing and regeneration device): Removes calcium and magnesium from the water to reduce water hardness.

– Protection Filter: Designed to remove particles in the range of 1 to 5 microns in diameter to protect the RO membrane.

– RO Model: Removal of ion, bacteria, pyrogen and so on.

– RO water supply: The constant pressure design inside the circulation pipe can ensure the safety and stability of HD treatment.

– Controller: Automatic system

– Disinfection device: Automatic disinfection control system, safe and easy to use.

Parameter and function

– Water quality: AAMI/ASAIO standard.

– High quality RO membrane (DOW), booster pump and controller.

– These automatic functions, including flushing after the machine is turned on, interval rinsing and disinfection, ensure convenience as there is no need for users to wait.

– Seamless membrane house design with no dead angle in circulation pipe effectively prevents bacteria growth.

– The first and second grades can be used separately to ensure normal water production during maintenance.

– Providing constant pressure in the soft water tank further ensures the stability of the device and the safety of water quality.

– These monitoring and alarm functions, including online water quality, water absence, pressure and voltage, enhance the safety of the device.

– Recycle concentrated water to save water.

– Sanitary materials, compact design and touch screen interface.

– Remote online monitoring (Optional function).

Technical specifications

Model Landing rate(L/H)25  Apply data (bed)
ROI-300/ROII-300 ≥300 6
ROI-500/ROII-500 ≥500 10
ROI-600/ROII-600 ≥600 12
ROI-750/ROII-750 ≥750 18
ROI-1000/ROII-1000 ≥1000 24
ROI-1250/ROII-1250 ≥1250 30
ROI-1500/ROII-1500 ≥1500 35
ROI-2000/ROII-2000 ≥2000 46
ROI-2500/ROII-2500 ≥2500 58

Technical Specifications